Total Organic Carbon Analyzer (TOC), Tekmar Apollo 9000, Refurbished
Apollo 9000 is comprised of the following:
• Syringe – for sample and reagent introduction
• Sparger – for inorganic analysis and sample preparation for TOC analysis
• Combustion Furnace
• Furnace Injection Port
• Water Trap – for moisture removal
• Permeation Tube – for moisture removal
• Corrosives Scrubber – for halogen removal
• NDIR Detector – for carbon dioxide detection
TOC Talk, the Apollo 9000 software, effortlessly monitors and performs calculations in a Windows 95/98/NT platform.
The following are some of the features of Apollo 9000 software, TOC Talk:
• Pre-defined Default Methods
• Multi-Point Calibration
• Outlier Deletion
• Wake-up and Shut-down
• Plotted Calibration Curve and Statistics
• Multiple Method Storage
• Analog Display of Peaks
• Continuous Scrolling of Results
• Diagnostics
Apollo 9000 TOC Analyzer technical specifications:
Chemistry:
• Oxidation by Combustion (from 680°C to 1000°C)
• IC by Acidification and Sparge Detector:
• Nondispersive Infrared (NDIR) Measurements:
• TOC (NPOC), TC-IC, TC, IC
Analytical:
• Range:
4 ppbC to 25,000 ppmC (sample volume, detector, and dilution dependent) using only 4 selectable analytical ranges
Precision:
• Typical 0.05% of full scale or
• RSD /- 3%, whichever is greater, over 3 replicates
Injection Volume:
• Up to 2 ml
Analysis Time:
• 1 to 3 minutes, typical
TOC Analysis Time:
• 15 minutes per triplicate, typical Liquid Handling:
• Syringe driver, 8-port valve
• Auto-dilution method provided from 4000 ppmC to 25,000 ppmC and/or difficult matrices
• Auto-rinsing with sample and rinse water
Sample Introduction:
• Automatic syringe injection
• Autosampler
• Solids Boat
• Manual Syringe
Controller:
• PC, Interface through Windows (95, 98 or NT), Password Protected
Data Handling:
• Automatic and customized Spreadsheet reports transferable to Microsoft Excel
• Real-time display of NDIR signal
• Outlier deletion
• Recalculations with different calibration curves
• Recalculations with different blank values
Calibration:
• Multi-point and automatic blank subtraction