Atomic Absorption Spectrophotometer (AAS), Persee AA990F, Flame

Atomic Absorption Spectrophotometer (AAS), Persee AA990F, Flame

Model No – Persee AA990F

Condition – New

Manufacturer – Persee

Shipping – 3 to 5 Business days (CAN/US)/ 5 to 10 Business days (International Shipping)

Lease Rental – Available

Add to cart

The AA900 Atomic Absorption Spectrometer is a versatile entry level instrument with a computer controlled Air/Acetylene flame for general laboratory requirements.

Due to its high sensitivity and excellent performance the instrument can be used for a wide range of applications.

 

The versatile instrument is available in three configurations:

  • AA990F-Flame atomizer only. Priced from $12,000 US
  • AA990G-Graphite furnace atomizer only. Priced from $15,000 US
  • AA990FG-Combined with flame and graphite furnace atomizer switch aromatically. Priced from $18,000 US

 

Common Applications include: Agricultural, Environmental, Petrochemical, Clinical.

 

Key Features

â—† Easy-to-use automation software

â—† Integrated safety precautions for user protection

â—† Precision accuracy with integrated correction system

â—† Life cycle value, best choice for less than 8 elements of analysis

 

Specifications

Model:Flame AA System-AA990F,Graphite AA System-AA990G,Combined Flame and Graphite Furnace AA System-AA990FG

Spectrometer System

Wavelength range 190-900 nm

Light source Hollow cathode lamp (HCL), Deuterium Arc lamp (D2)

Modulation Square wave pulse

Modulation frequency 100 Hz self reversal (SR) background, 400 Hz Deuterium Arc (D2) background

Monochromator

Grating 1800 grooves/mm diffraction grating

Blazed wavelength 250 nm

Focus 300 nm

Bandwidth 0.1, 0.2, 0.5, 1.0, 2.0, 5.0 nm

Scan mode Automatic

Photometric type Single beam

Wavelength accuracy ±0.15 nm

Wavelength resolution 0.2 nm±0.02 nm

Wavelength reproducibility <0.05 nm Spectral bandwidth 0.5, 1.0, 2.0, 5.0 nm Baseline stability 0.005 A/30 min Background correction Deuterium Arc (D2) 1.0 Abs, Self Reversal (SR) 3.0 Abs Flame Analysis Flame type Air/acetylene Sensitivity (Cu) 2 μg/ml Absorption>0.280 Abs

Characteristic Concentration Cu<0.02 μg/ml Detection limit Cu<0.004 μg/ml Repeatability Cu<0.7 % Burner head Titanium Alloy Nebulizer High–efficiency glass, acid proof available as an option Pre-mix chamber Corrosion resistant Atomizer selection Manual (AA990F) ,Automatic changeover (AA990FG) Safety features Burner switch, flame sensor, gas leak sensor, low gas pressure sensor, drain trap sensor, power loss protection Graphite Furnace Analysis Graphite head Transversally heated Temperature range Ambient-2650 ℃ Heating program Up to 10 steps. Drying, Ashing, Atomization, Cleaning Feedback Voltage and optical temperature control feedback Sensitivity (Cu) 50 ng/m Absorption>0.40 Abs

Detection limit Cd<0.004 ng/ml Repeatability Cu<2.0%, Cd<2.0% Graphite tubes Pyrolytically coated with platform Sample size Up to 20μl Graphite cooling Water circulator available Safety features Burner Switch, Flame Sensor, Gas leak Sensor, Drain Trap Sensor, Power Loss Protection, Circulation Water (graphite), Over Temperature Sensor (graphite) Data Processing Operating program AA Win software Analytical methods Flame AA, Flame AE, Graphite Furnace, Hydride Generation Readout Mode Continuous, manual, peak height, peak area Calibrations Multi-standard calibration, standard addition, interpolation Data storage Analytical results, instrument and measurement parameters, signal profile, calibration curve Power Requirements Main unit 220 VAC 50/60 Hz 200 W Graphite power supply 380 VAC 50/60 Hz instantaneous power 5kW current rating 7 kW Dimensions Main unit 44× 20 × 20 in 158 lb Graphite power supply 20 × 20 × 18 in 154 lb

SCIENTIFIC EQUIPMENT SOURCEGet Up To 50% Off

Subscribe to our weekly newsletter below and never miss the latest product or an exclusive offer.

Select your currency
CAD Canadian dollar