Signals RI range 1 to 1.75; measurement range

1/64 x 10 -5 DELTA RIU to 32 x 10 -5 DELTA RIU (full scale).

Noise < ± 2.5 x 10 -9 DELTA RIU peak-to-peak, flowing water at 1 ml/min, 0.8 second response time (10-90%)
Wander < ± 1 x 10 -8 DELTA RIU
Drift < 2.5 x 10 -7 DELTA RIU/hour, after warm-up.
Response Time 0.3, 0.8, or 5 seconds (10-90%).
Flow Cell 7 µl volume with 7 bar (103 psi) maximum pressure
Recommended pH range 2.3 to 9.5 standard. Solvents with a pH between 1.0 and 2.3 must not contain acids which attack stainless steel.
Temperature Control 30, 35, 40, 45, 50°C.
Display Single line, 4 1/2 digit display with real time display of output voltage (recorder or integrator
Analog Outputs One output integrator output 2 V maximum (32 mV/10 -5 DELTA RIU); two recorder outputs (0-1 mV and 0-10 mV).
Communications Four control inputs (marker, flush, balance not ready)
Environment Temperature range: 0 to 40°C
Humidity < 85% (non-condensing)
Power Requirements Line voltage: 100, 120, 220, 240 V AC (+5%, -10%)Line frequency: 48-66 Hz

Power consumption: 120 VA (410 BTU/hour)