HP 5890 Packed Injector, refurbished  

Interchangeable stainless steel liners/adapters, installed from inside the oven, are used with the packed column inlet, and with all detectors, depending upon the column to be installed. Liners for the packed column inlet are available in three sizes: one for 1/8•inchcolumns, one for 1/4•inchcolumns, and one for HP Series 530.

In addition, liners for the packed column inlet are available to accept glass inserts (discussed later) for reduced reactivity, to trap nonvolatile residues, or for use with an HP Series 530 u capillary column.

  • No liner is used with 1/4•inchpacked glass columns. The long leg of the column fits into the inlet body, replacing the liner. Packing and glass wool plug must be below the tip of the needle for best results.
  • Metal columns are installed with a liner appropriate for the column diameter.
  • If necessary, glass columns can be installed using a metal liner (preferably those accepting a glass insert), but this is not recommended. There may be problems with dead volume in connections, and preventing contact of sample with metal surfaces. Columns and Fittings
 

If you are a bargain hunter please give us a call, we always have instruments in stock, which have not been refurbished and can be sold as is in working condition.

Scientific Equipment Source provides service and training on site in most of central Ontario this includes neighboring cities Ajax, Whitby, Pickering, Oshawa, Toronto, Mississauga. We can also provide training and courses on site on chromatograph (GC), HPLC in the Durham region; this includes Pickering, Ajax, Whitby, Oshawa. For other customers we can be also flexible on arrangements. For the rest of customer any equipment for repair has to be sent back by a courier.